Installation

Selecting Positive or Negative-Edge Event Triggers

Figure 1-4

Positive/Negative-Edge EVENT TRIG jumpers

1. Remove the two screws that hold the E1330B board in place.

2. Remove the four screws that hold the 85330-60002 board in place.

3. Pull out the two boards at the same time. You can do this by pulling each

one out a little at a time, until they are both removed. Place the boards
on an anti-static surface.

Standard positive-edge operation is set at the factory by jumpering
E9 and E10 together.

4. To select negative-edge operation, move the jumper across E9 and E8.

5. Reinstall the boards and screws.