Runtime:source:source1:trigger, Default, Syntax – Agilent Technologies Multiple Channel Controller 85330A User Manual

Page 121: Example, Runtime:source:source2:trigger

Advertising
background image

85330A Multiple Channel Controller

7-11

Remote Programming Command Reference

Standard Command Reference

RUNTime:SOURce:SOURCE1:TRIGger

Selects IMM, TTL, REMote1, or REMote2 triggering for source 1 (the RF
source).

Default

The default trigger mode is TTL.

Syntax

RUNTime:SOURce:SOURCE1:TRIGger IMM | TTL

| REMote1 | REMote2;

Example

RUNT:SOUR:SOURCE1:TRIG IMM;

NOTE

In the fast source stepping mode (also known as Turbo), the trigger is set to
TTL. When using more than one multiple channel controller, the
CONTroller controller would set the source trigger parameter to remote
(REMote1 or REMote2), while the REMote1 or REMote2 controller sets its
parameter to TTL.

RUNTime:SOURce:SOURCE2:TRIGger

Selects IMM or TTL triggering for source 2 (the LO source).

Default

The default trigger mode is TTL.

Syntax

RUNTime:SOURce:SOURCE2:TRIGger IMM | TTL

| REMote1 | REMote2;

Example

RUNT:SOUR:SOURCE2:TRIG IMM;

NOTE

In the fast source stepping mode (also known as Turbo), the trigger is set to
TTL. When using more than one multiple channel controllers, the
CONTroller controller would set the source trigger parameter to remote
(REMote1 or REMote2), while the REMote1 or REMote2 controller sets its
parameter to TTL.

Advertising