Counter pulse delay test, Counter pulse delay test -9 – Agilent Technologies Multiple Channel Controller 85330A User Manual

Page 39

Advertising
background image

85330A Multiple Channel Controller

2-9

Performance Verification

Verifying the Multiple Channel Controller

Counter Pulse Delay
Test

The following test determines if the on-board counter is operating correctly.

1. Connect BNC cables between the SRC1 TRIG, RCVR READY

connectors and the oscilloscope as shown in

Figure 2-3 on page 2-10

.

2. On the oscilloscope press:

[RECALL] [1]

[TRIG] {Source 2} {Adjust: 1.5v} {

}

(+ E

dge}

[TIMEBASE] {1 us/Div} {Delay: 4 us} {Ref: Centr}

[SAVE] [2]

3. Press the following on the multiple channel controller:

[Clear Instr] [Select Instr] {85330A} {RESET} [More] {TEST} {COUNTER}

4. Press the following on the multiple channel controller:

[4] [Return]

5. The negative edge of the displayed pulse should be 4

±

1 microseconds.

To use the scope measurement feature, press:

[

t

v] {

t Marker: ON}.

Rotate the front panel knob until the vertical marker line intersects the
negative edge of the pulse. Note that the Start Marker reading will be the
pulse delay. See

Figure 2-4 on page 2-10

for more information.

6. On the oscilloscope press:

[TIME BASE] {1 us/Div} {Delay: 1000 us} [CLEAR DISPLAY]

7. Press the following on the multiple channel controller:

[1] [0] [0] [0] [Return]

8. The negative edge of the displayed pulse should be 1,000

±

1 microseconds.

To use the scope measurement feature, press:

[

t

v] {

t Marker: ON}.

Rotate the front panel knob until the vertical marker line intersects the
negative edge of the pulse. Note that the Start Marker reading will be the
pulse delay. See

Figure 2-4

for more information.

9. On the oscilloscope press:

[TIME BASE] {2 us/Div} {Delay: 15 ms} [CLEAR DISPLAY]

10. Press the following on the multiple channel controller:

[1] [5] [0] [0] [0] [Return]

11. The negative edge of the displayed pulse should be 15

± 0.

001

milliseconds.

To use the scope measurement feature, press:

[

t

v] {

t Marker: ON}.

Rotate the front panel knob until the vertical marker line intersects the

Advertising