Agilent Technologies Multiple Channel Controller 85330A User Manual

Page 84

Advertising
background image

6-10

85330A Multiple Channel Controller

Programming

To Use Direct Control

Other examples:

OUTPUT 70918;”ROUT:CLOS 2,(@103);”

Switch Port 2, SCU address 1, channel 3.

OUTPUT 70918;”ROUT:CLOS 1,(@2);”

Switch Port 1, SCU address 0, channel 2.

OUTPUT 70918,”ROUT:CLOS 1,(@132);”

Switch Port 1, SCU address 1, channel 32.

Sending multiple switch addresses is possible. In the following example:

An example is provided for standard systems (SCU address 0).

An example is provided for a typical custom system, with an SCU
address of 1.

Both examples select channel 1, 2 and 3 in sequence.

NOTE

Because the switches are SP2Ts or SP4Ts, when a channel on a module is
closed, all other channels on that module are open. For example, when 1 is
closed, 2, 3 and 4 are open; when 2 is closed, 1, 3, and 4 are open.

OUTPUT 70918;”ROUT:CLOS 1,(@1,2,3);”

example for standard systems

OUTPUT 70918;”ROUT:CLOS 1,(@101,102,103);”

example for custom systems

Another method of executing the above command is

OUTPUT 70918;”ROUT:CLOS 1,(@1:3);”

example for standard systems

OUTPUT 70918;”ROUT:CLOS 1,(@101:103);”

example for custom systems

The colon

:

represents 1 through 3 (or 101 through 103).

To set a delay between the closing of each switch state, the

ROUTe:DELAy

command is used. The input parameter is time in micro-seconds.

10 OUTPUT 70918;”*RST;”

Reset the 85330A.

20 OUTPUT 70918;”ROUT:DELA 10000;”

Set the time between switch states in micro-seconds.

30 OUTPUT 70918;”ROUT:CLOS 1,(@101:104);”

Close switch state 101 through 104.

Advertising