Runtime:timeout:remote, Default, Syntax – Agilent Technologies Multiple Channel Controller 85330A User Manual

Page 126: Example, Runtime:timeout:source, System

Advertising
background image

7-16

85330A Multiple Channel Controller

Remote Programming Command Reference

Standard Command Reference

RUNTime:TIMEout:REMote

Allows a timeout to occur if the REMote1 or REMote2 controller does not
respond in determined time.

Default

The default is 1 second (1000000 microseconds).

Syntax

RUNT:TIME:REM <microseconds>;

Example

RUNT:TIME:REM 1000000;

NOTE

This command is in revision A.02.31 or greater.

RUNTime:TIMEout:SOURce

Allows a timeout to occur if the source 1 or source 2 ready line does not
respond in determined time.

Default

The default is 1 second (1000000 microseconds).

Syntax

RUNT:TIME:SOUR <microseconds>;

Example

RUNT:TIME:SOUR 1000000;

SYSTem

Subsystem Command Syntax

SYSTem

:ERRor?