Ventilation requirements, Dimensions, Ventilation requirements -10 dimensions -10 – Agilent Technologies Multiple Channel Controller 85330A User Manual

Page 60

Advertising
background image

4-10

85330A Multiple Channel Controller

General Information

85330A Multiple Channel Controller Characteristics

Ventilation
Requirements

CAUTION

When installing the product in a cabinet, the convection into and out of the
product must not be restricted. The ambient temperature (outside the
cabinet) must be less than the maximum operating temperature of the
product by 4

°

C

for every 100 watts dissipated in the cabinet. If the total

power dissipated in the cabinet is greater than 800 watts, then forced
convection must be used.

Dimensions

Table 4-4

Size and Dimensions

Height without feet

177 mm

6.97 in

Height with feet

189 mm

7.44 in

Width

426 mm

16.75 in

Depth

510 mm

20.1 in

Depth with terminal blocks

569 mm

22.38 in

Net weight

11.8 kg

26 lbs