Runtime:initiate:imm, Default, Syntax – Agilent Technologies Multiple Channel Controller 85330A User Manual

Page 120: Example, Runtime:source:count, Runtime:initiate:imm -10 runtime:source:count -10

Advertising
background image

7-10

85330A Multiple Channel Controller

Remote Programming Command Reference

Standard Command Reference

RUNTime:INITiate:IMM

Initiates the run-time sequence and control of a measurement. All
parameters must be set up prior to initiating

RUNT:INIT:IMM.

Once the

measurement is initiated, it will run until completion or until a timeout has
occurred. If a timeout does occur, it will report the error to the error stack.

Default

The event trigger is defaulted to TTL. See other related commands for their
defaults.

Syntax

RUNTime:INITiate:IMM;

Example

RUNT:INIT:IMM;

RUNTime:SOURce:COUNt

Set the number of frequency points to be measured during the run-time
sequence. For each frequency point, each entry in the scan list will be
asserted and a trigger issued. If the SOURCE triggering method is IMM, no
source TTL trigger signal is issued and the 8530A is assumed to control the
sources. If the source triggering is TTL, a TTL signal is issued and the
run-time sequence waits for the source ready line.

Default

The default is 1.

Syntax

RUNTime:SOURce:COUNt <number of frequency points>;

Example

RUNT:SOUR:COUN 1;

RUNT:SOUR:COUN 51;

NOTE

For CW angle scan measurements the number of frequencies should be set
to 1.