85330a multiple channel controller characteristics, Environmental limits, Environmental limits -9 – Agilent Technologies Multiple Channel Controller 85330A User Manual

Page 59

Advertising
background image

85330A Multiple Channel Controller

4-9

General Information

85330A Multiple Channel Controller Characteristics

85330A Multiple Channel Controller

Characteristics

This product is designed for use in Installation Category II and Pollution
Degree 2 per IEC 1010 and 664 respectively.

CAUTION

Before switching on this instrument, make sure the line voltage selector
switch is set to the voltage of the mains supply and the correct fuse is
installed. Ensure the supply voltage is in the specified range.

Environmental Limits

Table 4-3

Environmental Conditions for the 85330A

Temperature:

For Operation

For Storage

+

0 to

+

55

°

C (32 to 131

°

F

)

40 to

+

70

°

C (

40 to 158

°

F

)

Humidity:

For Operation

5% to 95% at

+

40

°

C or less (non condensing)

5% to 95% at

+

65

°

C or less (non condensing)

Pressure Altitude:

For Operation or Storage

Less than 4,600 meters (15,000 feet)

Power Requirements:

Line Voltage
Fused at

115 or 230 Vac 50 to 400 Hz
3 A at 115 Vac
1.5 A at 230 Vac

Power Consumption:

85330A only
85330A

+

85331A (all off)

85330A

+

85332A (all off)

85330A

+ 2

85332As (all off)

1

78 VA
88 VA
92 VA
107 VA

1. This is the worst case for a standard system.