Selecting positive or negative-edge event triggers, Controller -8 – Agilent Technologies Multiple Channel Controller 85330A User Manual

Page 26

Advertising
background image

1-8

85330A Multiple Channel Controller

Installation

Selecting Positive or Negative-Edge Event Triggers

Viewing or Changing
the GPIB Address of
the Multiple Channel
Controller

The factory default setting of the GPIB address is 9. To view the GPIB
address:

Press

[Select Instr] {SYSTEM} {GPIB} {READ}

.

To change the current GPIB address:

1. Press

[Select Instr] {SYSTEM} {GPIB} {SET}

.

2. Enter the new address and press

[Return]

.

Selecting Positive or Negative-Edge Event

Triggers

By default, the positive-edge of the EVENT TRIG line starts the
measurement. To change this so a negative-edge starts the measurement,
perform the following procedure.

CAUTION

This device contains devices that are sensitive to static discharge. When you
remove the cover of this device, observe static safety precautions:

1. Place the unit on a grounded anti-static mat.

2. Wear a grounded wrist strap. Foot straps are only acceptable if you wear

one on each foot, and if you are standing on a grounded floor mat.

Refer to

Figure 1-4 on page 1-9

.