85330A Multiple Channel Controller

2-17

Performance Verification

AUX 1 and AUX 2

AUX 1 and AUX 2

Pulse Test

You can output one or more 500

µ

s pulses from AUX 1 or AUX 2 and

measure them with an oscilloscope. The number of pulses is selectable. The
pulses have a 50% duty cycle, and thus a total period of 1 ms. The amplitude
of the signal is roughly 0 to 4 volts. The pulses are not perfect square waves.
The leading edge rises higher than 4 volts, then curves down to
approximately 4 volts.

Measuring Pulses from
AUX 1 or AUX 2

1. Connect an oscilloscope to the AUX 1 or AUX.

2. Set the scope as necessary to view the waveform described above.

3. Press:

[Select Instr] {85330A}

{

}

{TEST}

{

}

{AUX 1}

or {

AUX 2}

4. The message

TEST AUXn OUTPUT: # OF TESTS

will appear.

5. Use the keyboard to enter the desired number of pulses. You can enter

the number in units or in scientific notation: For example:

100, or
1E2

6. Press the

[Return]

key. The pulses will appear on the scope.

NOTE

The 85330A will not perform any other functions during this test. If you
enter a large number, such as 1E6, it will take a while to finish. You can
abort the test by cycling line power.

NOTE

To test AUX 2, use the same basic procedure as shown above.