To Use Run-Time Control Mode

To Use Run-Time Control Mode

Setup of the
85330A Multiple
Channel Controller

The setup for run-time operation includes setting:

The event triggering method

The number of frequency points

The number of angular increments

List of switch states

Switch settling time

TTL trigger and ready timeouts

Using more than one controller

Once the 85330A is set up, send the

RUNT:INIT:IMM

command to begin run-time

operation.

Figure 6-2 on page 6-16

is a flow chart that illustrates the

run-time measurement sequence. Commands are described in

Chapter 7,

“Remote Programming Command Reference.”

Event Triggering

During run-time operation, the 85330A may use two different types of event
triggering, TTL or IMM. Most angle scan measurements will be set up for
TTL triggering, while TTL triggering can be suppressed for a frequency
response measurement, typically at a single angle. Also, this IMM mode
combined with the

*OPC?

command is useful for determining timing

information of a multiple-frequency, multi-parameter measurement.

Applicable command:

RUNTime:EVENte:TRIGger

Number of Frequency
Points

The 85330A does not require a start, stop, or frequency step size. In fact, it
does not require any frequency values. It only needs to know the number of
frequency points so it can issue the correct number of triggers.

Applicable command:

RUNTime:SOURce:COUNt

Number of Angular
Increments (Events)

The 85330A does not require any angular values. It only requires the number
of angular increments so it can issue the correct number of triggers.

Applicable command: RUNTime:EVENt:COUNt

List of Switch States

One must set up the list of channel addresses for the measurement. For each
event trigger received by the 85330A, each channel address entry will be
asserted and a subsequent trigger issued to begin a measurement of each data
point.

Applicable command:

RUNTime:SWITch:SCAN

Advertising