Rear Panel Features

The multiple channel controller may send several receiver trigger pulses
during the measurement sequence, depending on the instructions you sent to
it before starting the run-time mode.

RCVR READY

This positive-edge signal is sent by the receiver when it finished acquiring
data. This line connects to the RECEIVER READY line on the receiver.

AUX 1

Provides a user-controllable TTL line for special applications. The state of
AUX 1 must be set before entering the 85330A’s automatic run-time mode.
The state of AUX 1 is controlled with the ROUTe:CONTrol command.

AUX 2

Provides a user-controllable TTL line for special applications. The state of
AUX 2 can be set in two ways:

Before entering run-time mode using the ROUTe:CLOSe command.

During run-time using RUNTime:SWITch:SCAN command.

SRC 1 TRIG

Used in conjunction with SRC 1 READY, controls frequency switching in
the RF source much faster than is possible under 8530A control.This
connects to the TRIGGER IN line of the RF source.

SRC 1 READY

Used in conjunction with SRC 1 TRIG, controls frequency switching in the
RF source much faster than is possible under 8530A control. This connects
to the TRIGGER OUT line of the RF source.

SRC 2 TRIG

Used in conjunction with SRC 2 READY, controls frequency switching in
the LO source much faster than is possible under 8530A control. This
connects to the TRIGGER IN line of the LO source.

SRC 2 READY

Used in conjunction with SRC 2 TRIG, controls frequency switching in the
LO source much faster than is possible under 8530A control. This connects
to the TRIGGER OUT line of the LO source.

REMOTE 1 and 2

These connectors are used to add remote multiple channel controllers in
custom systems—they are not used with the standard instrument.