85330A Multiple Channel Controller

2-5

Performance Verification

Performance Verification Test

Network Analyzer

Turn ON the vector network analyzer. Adjust the controls for the following
setup:

Measurement Display

Display:

4 Channel

Channel 1:

S11

SWR

Channel 2:

S21

Log Mag

Channel 3:

S12

Log Mag

Channel 4:

S22

SWR

Domain:

Frequency

Start Frequency:

45 MHz

Stop Frequency:

40 GHz

Number of Points:

201

RF Source Power:

Maximum

Measurement:

Continuous

Averaging:

As required

Set up the rest of the measurement display for your convenience and ease of
viewing.

Cables

Connect one test port cable to Port 2 of the test set. The RF switch will be
connected directly to the test set Port 1. An example of this setup is shown in

Figure 2-1, “An Example Performance Verification Setup,” on page 2-3

.

Calibration

Perform a full, two port calibration at the test set Port 1, and at the end of the
test port cable connected to Port 2.

Save Setup

Save this instrument state setup in Save/Recall register 1. It will be used later
in the tests. Press:

[SAVE]

{1}