background image

85330A Multiple Channel Controller

7-9

Remote Programming Command Reference

Standard Command Reference

RUNTime:EVENt:TRIGger

The event trigger source allows either TTL triggering to control the run-time
sequencing or IMMediate triggering to occur.

TTL Mode

In TTL mode, the run-time control will:

1. Wait for an event trigger.

2. When the event trigger occurs, run-time control asserts each switch

address entry in the scan list (one at a time), issuing a measurement
trigger to the receiver for each.

3. Repeats step 2 for each frequency (defined using RUNT:SOUR:COUN).

Typically TTL mode is used for angle scans of CW or multi-frequency
responses.

IMM Mode

In IMM mode the 85330A issues the first trigger immediately on receiving
the INIT:IMM command and continues on with the sequence. Typically,
IMM mode is used for single angle frequency response measurements or
timing measurements.

Default

The default is TTL.

Syntax

RUNT:EVENt:TRIGger TTL | IMM;

Example

OUTPUT 70918;”RUNT:EVEN:TRIG TTL;”

OUTPUT 70918;”RUNT:EVEN:TRIG IMM;”