Using imm vs. ttl trigger for source 1 – Agilent Technologies Multiple Channel Controller 85330A User Manual

Page 92

Advertising
background image

6-18

85330A Multiple Channel Controller

Programming

To Use Run-Time Control Mode

Run-Time
Measurement
Sequence for Multiple
Controllers

The run-time measurement sequence for multiple controllers is similar to a
single controller shown earlier. See

“Starting Run-Time Mode” on page 6-14

for more information on this sequence. The sequence for multiple
controllers has additional steps to allow communications between the
controllers. Refer to

Figure 6-3 on page 6-19

and

Figure 6-4 on page 6-20

for the actual run-time measurement sequence.

Using IMM vs. TTL
Trigger for Source 1

Refer to “Source 1 IMM, TTL or REM1 trigger diamond in

Figure 6-3

.

When using multi frequencies:

The RF Source controlled by the master 85330A should use the TTL
trigger.

The other RF Source (not controlled by the master 85330A) should use
the IMM trigger.