Analog measurement wizard, Figure 73, Selecting input type and powering – Vaisala Hydromet Data Collection Platform User Manual

Page 156

Advertising
background image

User’s Guide ______________________________________________________________________

154 __________________________________________________________________M210785EN-E

Analog Measurement Wizard

Advanced user level of Lizard Setup Software offers the option of using
an analog measurement wizard. This measurement setup wizard was
created to make, for example, the interfacing of OEM measurement
instruments easier.

To configure an analog measurement with the measurement setup
wizard, proceed as follows:

1.

In the Equipment view, add the Generic 24 bit A/D component
and give it a more descriptive name. Connect it to the desired input.

2.

Proceed to the Measurements view and select the newly added
measurement. Rename the measurement component if necessary.

3.

Open the Utility menu and select the Wizard option. The window
presented in

Figure 73 on page 154

appears.

0711-087

Figure 73

Selecting Input Type and Powering

This window contains selections for:

-

Sensor output: voltage, current, or Pt100 resistance

-

Powering requirement from measurement channel's excitation
output. Note that the channels 0 to 3 can supply 12 VDC/
20 mA, while the channels 4 to 7 have only 1 mA or 100

μA

fixed current supply. If the device is powered by MAWS, a
further selection between the continuous and power saving
(device is powered only when measured) modes can be made.