The MEASure? and CONFigure Commands

See also “Measurement Configuration,” starting on page 51 in chapter 3.

For the range parameter,

MIN

selects the lowest range for the

selected function;

MAX

selects the highest range;

DEF

selects

autoranging.

For the resolution parameter, specify the resolution in the same units
as the measurement function, not in number of digits.

MIN

selects the

smallest value accepted, which gives the best resolution;

MAX

selects

the largest value accepted, which gives the least resolution;

DEF

selects the default resolution which is 5

1

2

digits slow (10

PLC

).

Note: You must specify a

range

to use the

resolution

parameter.

MEASure:VOLTage:DC? {<

range

>|MIN|MAX|DEF},{<

resolution

>|MIN|MAX|DEF}

Preset and make a dc voltage measurement with the specified range
and resolution. The reading is sent to the output buffer.

MEASure:VOLTage:DC:RATio? {<

range

>|MIN|MAX|DEF},{<

resolution

>|MIN|MAX|DEF}

Preset and make a dc:dc ratio measurement with the specified range
and resolution. The reading is sent to the output buffer. For ratio
measurements, the specified range applies to the signal connected to the

Input

terminals. Autoranging is automatically selected for reference

voltage measurements on the

Sense

terminals.

MEASure:VOLTage:AC? {<

range

>|MIN|MAX|DEF},{<

resolution

>|MIN|MAX|DEF}

Preset and make an ac voltage measurement with the specified range
and resolution. The reading is sent to the output buffer. For ac
measurements, resolution is actually fixed at 6

1

2

digits. The resolution

parameter only affects the front-panel display.

MEASure:CURRent:DC? {<

range

>|MIN|MAX|DEF},{<

resolution

>|MIN|MAX|DEF}

Preset and make a dc current measurement with the specified range
and resolution. The reading is sent to the output buffer.

4

Chapter 4 Remote Interface Reference
The MEASure? and CONFigure Commands