TRIGger:DELay {<

seconds

>|MINimum|MAXimum}

Insert a trigger delay between the trigger signal and each sample
that follows. If you do not specify a trigger delay, the multimeter
automatically selects a delay for you. Select from 0 to 3600 seconds.

MIN

= 0 seconds.

MAX

= 3600 seconds. [Stored in volatile memory]

TRIGger:DELay? [MINimum|MAXimum]

Query the trigger delay.

TRIGger:DELay:AUTO {OFF|ON}

Disable or enable an automatic trigger delay. The delay is determined
by function, range, integration time, and ac filter setting. Selecting a
specific trigger delay value automatically turns off the automatic trigger
delay. [Stored in volatile memory]

TRIGger:DELay:AUTO?

Query the automatic trigger delay setting. Returns “0” (

OFF

) or “1” (

ON

).

SAMPle:COUNt {<

value

>|MINimum|MAXimum}

Set the number of readings (samples) the multimeter takes per trigger.
Select from 1 to 50,000 readings per trigger.

MIN

= 1.

MAX

= 50,000.

[Stored in volatile memory]

SAMPle:COUNt? [MINimum|MAXimum]

Query the sample count.

TRIGger:COUNt {<

value

>|MINimum|MAXimum|INFinite}

Set the number of triggers the multimeter will accept before returning
to the “idle” state. Select from 1 to 50,000 triggers. The

INFinite

parameter instructs the multimeter to continuously accept triggers
(you must send a device clear to return to the “idle” state). Trigger count
is ignored while in local operation.

MIN

= 1.

MAX

= 50,000.

[Stored in volatile memory]

TRIGger:COUNt? [MINimum|MAXimum]

Query the trigger count. If you specify an infinite trigger count,
the query command returns “9.90000000E+37”.

4

Chapter 4 Remote Interface Reference
Triggering Commands

131