Boolean Parameters Boolean parameters represent a single binary
condition that is either true or false. For a false condition, the multimeter
will accept “

OFF

” or “0”. For a true condition, the multimeter will accept

ON

” or “1”. When you query a boolean setting, the instrument will

always return “0” or “1”. The following command uses a boolean parameter:

INPut:IMPedance:AUTO {OFF|ON}

String Parameters String parameters can contain virtually any set
of

ASCII

characters. A string must begin and end with matching quotes;

either with a single quote or with a double quote. You can include the
quote delimiter as part of the string by typing it twice without any
characters in between. The following command uses a string parameter:

DISPlay:TEXT <

quoted string

>

Output Data Formats

Output data will be in one of formats shown in the table below.

Type of Output Data

Non-reading queries
Single reading (IEEE-488)
Multiple readings (IEEE-488)
Single reading (RS-232)
Multiple readings (RS-232)

Output Data Format

< 80 ASCII character string
SD.DDDDDDDDESDD<nl>
SD.DDDDDDDDESDD,...,...,<nl>
SD.DDDDDDDDESDD<cr><nl>
SD.DDDDDDDDESDD,...,...,<cr><nl>

S

Negative sign or positive sign

D

Numeric digits

E

Exponent

<nl>

newline character

<cr>

carriage return character

4

Chapter 4 Remote Interface Reference
Output Data Formats

159