Using the Status Registers

The following example shows how you can use the multimeter’s status
registers to determine when a command sequence is completed. For
more information, see “The SCPI Status Model,” starting on page 134.
The example is shown in BASIC and QuickBASIC (see page 190).

10 REAL Aver,Min_rdg,Max_rdg

20 INTEGER Val,Hpib,Mask,Task

30 ASSIGN @Dmm TO 722

40 CLEAR 7

! Clear GPIB and dmm

50 OUTPUT @Dmm; "*RST"

! Reset dmm

60 OUTPUT @Dmm; "*CLS"

! Clear dmm status registers

70 OUTPUT @Dmm; "*ESE 1" ! Enable "operation complete" bit to set

! "standard event" bit in status byte

80 OUTPUT @Dmm; "*SRE 32" ! Enable "standard event" bit in status byte

! to pull the IEEE-488 SRQ line

90 OUTPUT @Dmm; "*OPC?"

! Assure synchronization

100 ENTER @Dmm; Val

110 !

120 ! Configure the multimeter to make measurements

130 !

140 OUTPUT @Dmm; "CONF:VOLT:DC 10" ! Set dmm to 10 volt dc range

150 OUTPUT @Dmm; "VOLT:DC:NPLC 10" ! Set the integration time to 10 PLCs

160 OUTPUT @Dmm; "TRIG:COUN 100" ! Dmm will accept 100 triggers

170 OUTPUT @Dmm; "CALC:FUNC AVER;STAT ON" ! Select min-max and enable math

180 OUTPUT @Dmm; "INIT"

! Place dmm in "wait-for-trigger" state

190 OUTPUT @Dmm; "*OPC"

! Set "operation complete" bit in standard event

! registers when measurement is complete

200 !

210 Hpib=7

220 ON INTR Hpib GOSUB Read_data

230 Mask=2

! Bit 1 is SRQ

240 ENABLE INTR Hpib;Mask ! Enable SRQ to interrupt the program

250 !

260 ! Execute other tasks while waiting for data

270 !

GPIB Operation Using BASIC

Continued on next page

>

Chapter 6 Application Programs
Using the Status Registers

188