Remote Interface Selection

The multimeter is shipped with both an

GPIB

(

IEEE-488

) interface

and an

RS-232

interface. Only one interface can be enabled at a time.

The

GPIB

interface is selected when the multimeter is shipped from

the factory.

The remote interface can be set only from the front-panel.

The interface selection is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.

If you select the

GPIB

interface, you must select a unique address for

the multimeter. The

GPIB

address is displayed when you turn on the

multimeter.

If you select the

RS-232

interface, you must set the baud rate and

parity for the multimeter. “

RS-232

” is displayed when you turn on the

multimeter.

If you select the

RS-232

interface and then set the

GPIB

address to the

talk only address (31), the multimeter will send readings over the

RS-232

interface when in the local mode.

There are certain restrictions to be aware of when you are selecting
the remote interface (see also “Programming Language Selection,” on
page 94). The only programming language supported on

RS-232

is

SCPI

.

Front-Panel Operation:

2: INTERFACE (I/O MENU)

See also “To Select the Remote Interface,” on page 162.

SCPI Language
3478A Language
Fluke 8840A Language

GPIB / 488

X
X

X

RS-232

X
Not allowed
Not allowed