-151

Invalid string data
An invalid character string was received. Check to see if you have
enclosed the character string in single or double quotes and that the
string contains valid

ASCII

characters. Example:

DISP:TEXT ’ON

(the ending quote is missing).

-158

String data not allowed
A character string was received but is not allowed for the command.
Check the list of parameters to verify that you have used a valid
parameter type. Example:

CALC:STAT ’ON’

-160 to -168

Block data errors
The multimeter does not accept block data.

-170 to -178

Expression errors
The multimeter does not accept mathematical expressions.

-211

Trigger ignored
A Group Execute Trigger (

GET

) or

*TRG

was received but the trigger

was ignored. Make sure the multimeter is in the “wait-for-trigger” state
before issuing a trigger, and make sure the correct trigger source is selected.

-213

Init ignored
An

INITiate

command was received but could not be executed because

a measurement was already in progress. Send a device clear to halt a
measurement in progress and place the multimeter in the “idle” state.

-214

Trigger deadlock
A trigger deadlock occurs when the trigger source is

BUS

and a

READ?

command is received.

5

Chapter 5 Error Messages
Execution Errors

175