RS-232 Operation Using QuickBASIC

The following example shows how to send command instructions and
receive command responses over the

RS-232

interface using QuickBASIC.

CLS

LOCATE 1, 1

DIM cmd$(100), resp$(1000)

’ Set up serial port for 9600 baud, even parity, 7 bits;

’ Ignore Request to Send and Carrier Detect; Send line feed,

’ enable parity check, reserve 1000 bytes for input buffer

OPEN "com1:9600,e,7,2,rs,cd,lf,pe" FOR RANDOM AS #1 LEN = 1000

’ Put the multimeter into the remote operation mode

PRINT #1, ":SYST:REM"

’ Query the multimeter’s id string

PRINT #1, "*IDN?"

LINE INPUT #1, resp$

PRINT "*IDN? returned: ", resp$

’ Ask what revision of SCPI the multimeter conforms to

PRINT #1, ":SYST:VERS?"

LINE INPUT #1, resp$

PRINT ":SYST:VERS? returned: ", resp$

’ Send a message to the multimeter’s display, and generate a beep

PRINT #1, ":SYST:BEEP;:DISP:TEXT ’34401A’"

’ Configure the multimeter for dc voltage readings,

’ 10 V range, 0.1 V resolution, 4 readings

PRINT #1, ":CONF:VOLT:DC 10,0.1;:SAMP:COUN 4"

’ Trigger the readings, and fetch the results

PRINT #1, ":READ?"

LINE INPUT #1, resp$

PRINT ":READ? returned: ", resp$

END

RS-232 Operation Using QuickBASIC

Chapter 6 Application Programs
RS-232 Operation Using QuickBASIC