Agilent Technologies 34401A User Manual

Page 187

Advertising
background image

703

Invalid secure code
An invalid calibration security code was received when attempting to
unsecure or secure the multimeter. You must use the same security code
to unsecure the multimeter as was used to secure it, and vice versa. The
security code may contain up to 12 alphanumeric characters. The first
character must be a letter.

704

Secure code too long
A security code was received which contained more than 12 characters.

705

Cal aborted
A calibration in progress is aborted when you press any front-panel key,
send a device clear, or change the local/remote state of the multimeter.

706

Cal value out of range
The specified calibration value (

CALibration:VALue

) is invalid for the

present function and range.

707

Cal signal measurement out of range
The specified calibration value (

CALibration:VALue

) does not match

the signal applied to the multimeter.

708

Cal signal frequency out of range
The input signal frequency for an ac calibration does not match the
required input frequency for calibration.

709

No cal for this function or range
You cannot perform calibrations for ac current, period, continuity, diode,
ratio, or on the 100 M

range.

710

Full scale correction out of range

720

Cal DCV offset out of range

721

Cal DCI offset out of range

722

Cal RES offset out of range

723

Cal FRES offset out of range

724

Extended resistance self cal failed

5

Chapter 5 Error Messages