RS-232 Interface Commands

Use the front-panel I/O MENU to select the baud rate, parity, and
number of data bits (see pages 163 and 164 for more information).


SYSTem:LOCal

Place the multimeter in the local mode for

RS

-232 operation. All keys on

the front panel are fully functional.

SYSTem:REMote

Place the multimeter in the remote mode for

RS

-232 operation.

All keys on the front panel, except the

LOCAL

key, are disabled.

It is very important that you send the

SYSTem:REMote

command

to place the multimeter in the remote mode. Sending or receiving data
over the RS-232 interface when not configured for remote operation can
cause unpredictable results.

SYSTem:RWLock

Place the multimeter in the remote mode for

RS

-232 operation.

This command is the same as the

SYSTem:REMote

command except

that all keys on the front panel are disabled, including the

LOCAL

key.

Ctrl-C

Clear the operation in progress over the

RS

-232 interface and discard

any pending output data. This is equivalent to the IEEE-488 device clear
action over the GPIB interface.

4

Chapter 4 Remote Interface Reference
RS-232 Interface Commands

153