Agilent Technologies 34401A User Manual

Page 184

Advertising
background image

501

Isolator UART framing error

502

Isolator UART overrun error

511

RS-232 framing error

512

RS-232 overrun error

513

RS-232 parity error

514

Command allowed only with RS-232
There are three commands which are only allowed with the

RS-232

interface:

SYSTem:LOCal

,

SYSTem:REMote

, and

SYSTem:RWLock

.

521

Input buffer overflow

522

Output buffer overflow

531

Insufficient memory
There is not enough memory to store the requested number of readings
in internal memory using the

INITiate

command. The product of the

sample count (

SAMPle:COUNt

) and the trigger count (

TRIGger:COUNt

)

must not exceed 512 readings.

532

Cannot achieve requested resolution
The multimeter cannot achieve the requested measurement resolution.
You may have specified an invalid resolution in the

CONFigure

or

MEASure

command.

540

Cannot use overload as math reference
The multimeter cannot store an overload reading (9.90000000E+37) as
the math reference for null or dB measurements. The math state is
turned off as a result of this condition.

550

Command not allowed in local
The multimeter received a

READ?

or MEASure? command while in the

local mode. During

RS-232

operation, you should always execute the

SYSTem:REMote

command before sending other commands over the

interface.

Chapter 5 Error Messages
Execution Errors

178