Agilent Technologies 34401A User Manual

Page 80

Advertising
background image

External Triggering In the external trigger mode, the multimeter
will accept a hardware trigger applied to the Ext Trig terminal.
The multimeter takes one reading, or the specified number of readings
(sample count), each time Ext Trig receives a low-true pulse.

See also “External Trigger Terminal,” on page 83.

The multimeter buffers one external trigger. This means that if the
multimeter is taking a reading and another external trigger occurs,
that trigger is accepted (a “Trigger ignored” error is not reported).
After the reading in progress is complete, the stored trigger satisfies
the trigger source and then the trigger is issued.

Front-Panel Operation: The external trigger mode is like the single
trigger mode except that you apply the trigger to the Ext Trig
terminal. Pressing

Single

to enable the single trigger mode also

enables the external trigger mode. The

Trig

annunciator turns on

when the multimeter is waiting for an external trigger.

The front-panel

Single

key is disabled when in remote.

Remote Interface Operation:

TRIGger:SOURce EXTernal

Chapter 3 Features and Functions
Triggering

74