Measurement Configuration Commands

See also “Measurement Configuration,” starting on page 51 in chapter 3.

FUNCtion "<

function

>"

Select a measurement function. The function must be enclosed in quotes
in the command string (

FUNC "VOLT:DC"

). Specify one of the following

strings.

FUNCtion?

Query the measurement function and return a quoted string.

<

function

>:RANGe {<

range

>|MINimum|MAXimum}

Select the range for the selected function. For frequency and period
measurements, ranging applies to the signal’s input voltage, not its
frequency (use

FREQuency:VOLTage

or

PERiod:VOLTage

).

MIN

selects

the lowest range for the selected function.

MAX

selects the highest

range. [Stored in volatile memory]

<

function

>:RANGe? [MINimum|MAXimum]

Query the range for the selected function.

<

function

>:RANGe:AUTO {OFF|ON}

Disable or enable autoranging for the selected function. For frequency
and period, use

FREQuency:VOLTage

or

PERiod:VOLTage

.

Autorange thresholds: Down range at <10% of range; Up range
at >120% of range. [Stored in volatile memory]

<

function

>:RANGe:AUTO?

Query the autorange setting. Returns “0” (

OFF

) or “1” (

ON

).

VOLTage:DC
VOLTage:DC:RATio
VOLTage:AC
CURRent:DC
CURRent:AC
RESistance (

2-wire ohms

)

FRESistance (

4-wire ohms

)

FREQuency
PERiod
CONTinuity
DIODe

4

Chapter 4 Remote Interface Reference
Measurement Configuration Commands