To set the parity – Agilent Technologies 34401A User Manual

Page 170

Advertising
background image

To Set the Parity

You can select the parity for

RS-232

operation. The multimeter is

configured for even parity with 7 data bits when shipped from the
factory. See also “Parity Selection,” on page 93.

1 Turn on the front-panel menu.

A: MEAS MENU

2 Move across to the

I/O MENU

choice on this level.

E: I/O MENU

3 Move down a level and then across to the

PARITY

command.

4: PARITY

4 Move down to the “parameter” level to select the parity.

Use the left/right arrow keys to see the parity choices. Choose from one of
the following: None (8 data bits), Even (7 data bits), or Odd (7 data bits).
When you set parity, you are indirectly setting the number of data bits.

EVEN: 7 BITS

5 Save the change and turn off the menu.

The parity selection is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.

< <

Shift <

On/Off

Auto/Man

ENTER

< <

Chapter 4 Remote Interface Reference
To Set the Parity

164