Display Control

To speed up your measurement rate, or for security reasons, you may
want to turn off the front-panel display. From the remote interface, you
can also display a 12-character message on the front panel.

When the display is turned off, readings are not sent to the display
and all display annunciators except

ERROR

and

Shift

are disabled.

Front-panel operation is otherwise unaffected by turning off the display.

The display state is stored in volatile memory; the display is enabled
when power has been off or after a remote interface reset.

You can display a message on the front panel by sending a
command from the remote interface. The multimeter can display up
to 12 characters of the message on the front panel; any additional
characters are truncated. Commas, periods, and semicolons share a
display space with the preceding character, and are not considered
individual characters. When a message is displayed, readings are not
sent to the display.

Sending a message to the display from the remote interface overrides
the display state; this means that you can display a message even if
the display is turned off.

Front-Panel Operation:

5: DISPLAY (SYS MENU)

The display always turns on for menu operation; this means that
even when the display is turned off, you can still operate the menu.

Remote Interface Operation:

DISPlay {OFF|ON}

disable/enable the

display

DISPlay:TEXT <

quoted string

>

display the string enclosed in quotes

DISPlay:TEXT:CLEar

clear the displayed message

The following command string shows how to display a message on the
front panel.

"DISP:TEXT ’HELLO’"

3

Chapter 3 Features and Functions
System-Related Operations