To make dbm measurements – Agilent Technologies 34401A User Manual

Page 47

Advertising
background image

To Make dBm Measurements

The dBm operation calculates the power delivered to a resistance
referenced to 1 milliwatt.

dBm = 10

×

Log

10

( reading

2

/

reference resistance

/

1 mW )

Select

DC V

or

AC V

.

The factory setting for the reference resistance is 600

. To select a

different value, press

Shift >

(Menu Recall) after enabling dBm

operations. This takes you to the “

dBm REF R

” command in the

MATH MENU

(only if dBm is enabled).

Go down to the “parameter” level, and then select a value: 50, 75,
93, 110, 124, 125, 135, 150, 250, 300, 500, 600, 800, 900, 1000,
1200, or 8000 ohms.

The reference resistance is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.

Enables dBm operation;
Press again to disable.

Math annunciator is on when
dBm operation is enabled.

To read / edit the dBm reference resistance,
use the MATH menu.