Front-Panel Operation (continued)

To set the delay to 0 seconds, select the “parameter” level of the

TRIG

DELAY

command. Move the flashing cursor to the “units” location on

the right side of the display. Press

until

ZERO DELAY

is reached,

then press Menu Enter.

To select the automatic trigger delay, select the “parameter” level
of the

TRIG DELAY

command. Move the flashing cursor to the

“units” location on the right side of the display. Press

until

AUTO DELAY

is reached, then press Menu Enter.

Remote Interface Operation:

You can use the following command to set the trigger delay.

TRIGger:DELay {<

seconds

>|MINimum|MAXimum}

You can use the following command to set an automatic trigger delay.

TRIGger:DELay:AUTO {OFF|ON}

Trigger Delay

(continued)

ZERO DELAY

AUTO DELAY

Chapter 3 Features and Functions
Triggering

80