Remote Interface Operation: You can use the following commands to
make dB measurements. Math must be enabled before you can store a
value to the Relative Register.

CALCulate:FUNCtion DB
CALCulate:STATe {OFF|ON}
CALCulate:DB:REFerence {<

value

>|MINimum|MAXimum}

dBm Measurements

The dBm operation calculates the power delivered to a resistance
referenced to 1 milliwatt.

dBm = 10

×

Log

10

( reading

2

/

reference resistance

/

1 mW )

Applies to dc voltage and ac voltage measurements only.

You can choose from 17 different reference resistance values.
The factory setting for the reference resistance is 600

.

The choices are: 50, 75, 93, 110, 124, 125, 135, 150, 250, 300, 500,
600, 800, 900, 1000, 1200, or 8000 ohms.

The reference resistance is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.

Front-Panel Operation: After enabling dBm, you can select a new
reference resistance by pressing

Shift >

(Menu Recall). Turning

on the menu does not disable the dBm operation; the multimeter
will resume taking measurements when you turn off the menu.

4: dBm REF R (MATH MENU)

See also “To Make dBm Measurements,” on page 41.