Frequency and period characteristics – Agilent Technologies 34401A User Manual

Page 226

Advertising
background image

Frequency and Period Characteristics

Frequency and Period Characteristics

Accuracy Specifications

±

( % of reading ) [ 1 ]

Function Range

[ 3 ]

Frequency

24 Hour

[ 2 ]

23°C

±

1°C

1 Year
23°C

±

5°C

Temperature
Coefficient/°C
0°C – 18°C
28°C – 55°C

Frequency,
Period
[ 4 ]

100 mV
to
750 V

3 Hz 5 Hz
5 Hz 10 Hz
10 Hz 40 Hz
40 Hz 300 kHz

0.10
0.05
0.03
0.006

0.10
0.05
0.03
0.01

0.10
0.05
0.03
0.01

0.005

0.005

0.001

0.001

Additional Low-Frequency Errors ( % of reading )

[ 4 ]

Frequency

Resolution
6

1

2

5

1

2

4

1

2

0 0.12

0.12

0 0.17

0.17

0 0.2

0.2

0 0.06 0.21
0 0.03

0.21

0 0.01 0.07
0 0 0.02

3 Hz 5 Hz
5 Hz 10 Hz
10 Hz 40 Hz
40 Hz 100 Hz
100 Hz 300 Hz
300 Hz 1 kHz
> 1 kHz

Transfer Accuracy ( typical )

0.0005% of reading

Conditions:

Within 10 minutes and

±

0.5°C.

Within

±

10% of initial value.

Following a 2-hour warm-up.

For inputs > 1 kHz and > 100 mV.

Using 6

1

2

digit slow resolution ( 1 second gate time ).

Measurements are made using accepted metrology practices.

90 Day
23°C

±

5°C

Chapter 8 Specifications

220

Advertising