Remote Interface Operation: You can use the following commands to
make min-max measurements.

CALCulate:FUNCtion AVERage
CALCulate:STATe {OFF|ON}
CALCulate:AVERage:MINimum?

read the minimum value

CALCulate:AVERage:MAXimum?

read the maximum value

CALCulate:AVERage:AVERage?

read the average of all readings

CALCulate:AVERage:COUNt?

read the count

A new command is available starting with firmware Revision 2 which
allows you to take readings using

INITiate

without storing them in

internal memory. This command may be useful with the min-max
operation since it allows you to determine the average of a series of
readings without storing the individual values.

DATA:FEED RDG_STORE, ""

do not store readings

DATA:FEED RDG_STORE, "CALCulate"

store readings (default)

See page 126 for more information on using the

DATA:FEED

command.

Null (Relative) Operation

When making null measurements, also called relative, each reading
is the difference between a stored null value and the input signal.
One possible application is in making more accurate two-wire ohms
measurements by nulling the test lead resistance.

Result = reading null value

Applies to all measurement functions, except continuity, diode, and ratio.

The null value is adjustable and you can set it to any value between
0 and

±

120% of the highest range, for the present function.

The null value is stored in volatile memory; the value is cleared
when power has been off, after a remote interface reset, or after a
function change.

3