Automatic Trigger Delays

If you do not specify a trigger delay, the multimeter selects an
automatic delay for you. The delay is determined by function, range,
integration time, and ac filter setting.

DC Voltage and DC Current (for all ranges):

Resistance (2-wire and 4-wire):

AC Voltage and AC Current (for all ranges):

Frequency and Period:

Integration Time

NPLC

1

NPLC

<

1

Trigger Delay

1.5

ms

1.0

ms

Range

100

1 k

10 k

100 k

1 M

10 M

100 M

Trigger Delay
(For NPLC

1)

1.5 ms

1.5

ms

1.5

ms

1.5

ms

15

ms

100

ms

100

ms

AC Filter

Slow
Medium
Fast

Trigger Delay

7.0

sec

1.0

sec

600

ms

Remote or single/external trigger

Range

100

1 k

10 k

100 k

1 M

10 M

100 M

Trigger Delay
(For NPLC < 1)

1.0

ms

1.0

ms

1.0

ms

1.0

ms

10

ms

100

ms

100

ms

AC Filter

Slow
Medium
Fast

Trigger Delay

1.5

sec

200

ms

100

ms

Front panel with auto trigger ON

Trigger Delay

1.0 sec

Remote or single/external trigger


Trigger Delay

0 sec

Front panel with auto trigger ON

3