DC Input Resistance

Normally, the multimeter’s input resistance is fixed at 10 M

for all

dc voltage ranges to minimize noise pickup. To reduce the effects of
measurement loading errors, you can set the input resistance to greater
than 10 G

for the 100 mVdc, 1 Vdc, and 10 Vdc ranges.

Applies to dc voltage measurements and is disabled for all other functions.

The input resistance setting is stored in volatile memory; the
multimeter selects 10 M

(for all dc voltage ranges) when power

has been off or after a remote interface reset.

Front-Panel Operation: Select from the menu the 10 M

mode (fixed

resistance for all dc voltage ranges) or the >10 G

mode. The default

is 10 M

.

3: INPUT R (MEAS MENU)

Remote Interface Operation: You can enable or disable the automatic
input resistance mode. With

AUTO OFF

(default), the input resistance

is fixed at 10 M

for all ranges. With

AUTO ON

, the input resistance

is set to >10 G

for the three lowest dc voltage ranges. The

CONFigure

and

MEASure?

commands automatically turn

AUTO OFF

.

INPut:IMPedance:AUTO {OFF|ON}

Input Resistance

Input Resistance

100mV, 1V, 10V ranges

100V, 1000V ranges

Fixed Resistance ON

(default)

10 M

10 M

Fixed Resistance OFF

> 10 G

10 M

3

Chapter 3 Features and Functions