Frequency and Period Characteristics

Frequency and Period
Measurement Method:

Voltage Ranges:

Gate Time:

Reciprocal-counting technique.
AC-coupled input using the
ac voltage measurement function.
100 mV rms full scale to 750 V rms.
Auto or manual ranging.
10 ms, 100 ms, or 1 sec

Settling Considerations

Errors will occur when attempting to measure the frequency or
period of an input following a dc offset voltage change. The input
blocking RC time constant must be allowed to fully settle ( up to
1 sec ) before the most accurate measurements are possible.

Measurement Considerations

All frequency counters are susceptible to error when
measuring low-voltage, low-frequency signals. Shielding
inputs from external noise pickup is critical for minimizing
measurement errors.

Measuring Characteristics

Function
Frequency,
Period

6

1

2

5

1

2

4

1

2

1
9.8
80

System Speeds

[ 5 ]

Configuration Rates
Autorange Time
ASCII readings to RS-232
ASCII readings to GPIB
Max. Internal Trigger Rate
Max. External Trigger Rate to Memory
Max. External Trigger Rate to GPIB/RS-232

14/sec
<0.6 sec
55/sec
80/sec
80/sec
80/sec
80/sec

[ 1 ]

Specifications are for 1-hour warm-up at 6

1

2

digits.

[ 2 ]

Relative to calibration standards.

[ 3 ]

20% overrange on all ranges, except 750 Vac range.

[ 4 ]

Input > 100 mV.

For 10 mV to 100 mV inputs, multiply % of reading
error x10.

[ 5 ]

Speeds are for 4

1

2

digits, Delay 0, Display OFF,

and Fast AC filter.

Operating Characteristics

[ 5 ]

Digits

Reading/s

8

Chapter 8 Specifications