Status Reporting Commands

SYSTem:ERRor?

Query the multimeter’s error queue. Up to 20 errors can be stored in the
queue. Errors are retrieved in first-in-first out (

FIFO

) order. Each error

string may contain up to 80 characters.

STATus:QUEStionable:ENABle <

enable value

>

Enable bits in the Questionable Data enable register. The selected bits
are then reported to the Status Byte.

STATus:QUEStionable:ENABle?

Query the Questionable Data enable register. The multimeter returns a
binary-weighted decimal representing the bits set in the enable register.

STATus:QUEStionable:EVENt?

Query the Questionable Data event register. The multimeter returns a
decimal value which corresponds to the binary-weighted sum of all bits
set in the register.

STATus:PRESet

Clear all bits in the Questionable Data enable register.

*CLS

Clear the Status Byte summary register and all event registers.

*ESE <

enable value

>

Enable bits in the Standard Event enable register. The selected bits are
then reported to the Status Byte.

*ESE?

Query the Standard Event enable register. The multimeter returns a
decimal value which corresponds to the binary-weighted sum of all bits
set in the register.

Chapter 4 Remote Interface Reference
Status Reporting Commands