Autozero – Agilent Technologies 34401A User Manual

Page 65

Advertising
background image

Autozero

When autozero is enabled (default), the multimeter internally
disconnects the input signal following each measurement, and takes a
zero reading. It then subtracts the zero reading from the preceding
reading. This prevents offset voltages present on the multimeter’s input
circuitry from affecting measurement accuracy.

When autozero is disabled, the multimeter takes one zero reading and
subtracts it from all subsequent measurements. It takes a new zero
reading each time you change the function, range, or integration time.

Applies to dc voltage, dc current, and 2-wire ohms measurements only.
Autozero is enabled when you select 4-wire ohms or ratio measurements.

The autozero mode is stored in volatile memory; the multimeter
automatically enables autozero when power has been off or after a
remote interface reset.

Front-Panel Operation: The autozero mode is set indirectly when you
set the resolution.

Remote Interface Operation: The

OFF

and

ONCE

parameters have a

similar effect. Autozero

OFF

does not issue a new zero measurement.

Autozero

ONCE

issues an immediate zero measurement.

ZERO:AUTO {OFF|ONCE|ON}

Resolution Choices

Fast 4 Digit

*

Slow 4 Digit

Fast 5 Digit

*

Slow 5 Digit

(default)

*

Fast 6 Digit

Slow 6 Digit

Integration Time

0.02 PLC
1 PLC

0.2 PLC
10 PLC

10 PLC
100 PLC

Autozero

Off

On


Off

On

On

On

* These settings configure the multimeter just as if you had
pressed the corresponding “DIGITS” keys from the front panel.

3

Chapter 3 Features and Functions
Measurement Configuration

59

Advertising