Dc characteristics – Agilent Technologies 34401A User Manual

Page 222

Advertising
background image

DC Characteristics

DC Characteristics

Accuracy Specifications

±

( % of reading + % of range ) [ 1 ]

Function

Range [ 3 ]

Test Current or
Burden Voltage

90 Day
23°C

±

5°C

1 Year
23°C

±

5°C

Temperature
Coefficient /°C
0°C – 18°C
28°C – 55°C

DC Voltage

100.0000 mV
1.000000 V
10.00000 V
100.0000 V
1000.000 V

0.0030 + 0.0030
0.0020 + 0.0006
0.0015 + 0.0004
0.0020 + 0.0006
0.0020 + 0.0006

0.0040 + 0.0035
0.0030 + 0.0007
0.0020 + 0.0005
0.0035 + 0.0006
0.0035 + 0.0010

0.0050 + 0.0035
0.0040 + 0.0007
0.0035 + 0.0005
0.0045 + 0.0006
0.0045 + 0.0010

0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001

Resistance
[ 4 ]

100.0000

1.000000 k

10.00000 k

100.0000 k

1.000000 M

10.00000 M

100.0000 M

1 mA
1 mA
100

µ

A

10

µ

A

5

µ

A

500 nA
500 nA || 10 M

0.0030 + 0.0030
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0005
0.002 + 0.001
0.015 + 0.001
0.300 + 0.010

0.008 + 0.004
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.020 + 0.001
0.800 + 0.010

0.010 + 0.004
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.040 + 0.001
0.800 + 0.010

0.0006 + 0.0005
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0004
0.1500 + 0.0002

DC Current

10.00000 mA
100.0000 mA
1.000000 A
3.000000 A

< 0.1 V
< 0.6 V
< 1 V
< 2 V

0.005 + 0.010
0.01 + 0.004
0.05 + 0.006
0.10 + 0.020

0.030 + 0.020
0.030 + 0.005
0.080 + 0.010
0.120 + 0.020

0.050 + 0.020
0.050 + 0.005
0.100 + 0.010
0.120 + 0.020

0.002 + 0.0020
0.002 + 0.0005
0.005 + 0.0010
0.005 + 0.0020

Continuity

1000.0

1 mA

0.002 + 0.030

0.008 + 0.030

0.010 + 0.030

0.001 + 0.002

Diode Test

[12]

1.0000 V

1 mA

0.002 + 0.010

0.008 + 0.020

0.010 + 0.020

0.001 + 0.002

DC:DC

Ratio

100 mV
to
1000 V

( Input Accuracy ) + ( Reference Accuracy )

Input Accuracy = accuracy specification for the HI-LO input signal.
Reference Accuracy = accuracy specification for the HI-LO reference input signal.

Transfer Accuracy ( typical )

(

24 hour %

of

range

error

)

2

Conditions:

Within 10 minutes and

±

0.5°C.

Within

±

10% of initial value.

Following a 2-hour warm-up.

Fixed range between 10% and 100% of full scale.

Using 6

1

2

digit slow resolution ( 100 PLC ).

Measurements are made using accepted metrology practices.

24 Hour [ 2 ]
23°C

±

1°C

Chapter 8 Specifications

216

Advertising