7 Methods

IC Net 2.2

144

To reset the

Run

number press [Ctrl] [F8].

The

Analysis number

window is opened

where the run number for the next run can
be modified to the desired value.

Batch

X/Y: total number of started sample queues
/ current analysis number in the sample
queue (read-only).


7.2.2 Sample

Tab

Sample

of the passport with sample information.

Info 1

First sample description (max. 256 charac-
ters).

Info 2

Second sample description (max. 256
characters).

Volume

Injected volume in

µ

L.

Dilution

Dilution of the sample.

Vial number

Autosampler vial position to take sample
from.

Amount

Sample amount. If this value is different for
the calibration run (c) and the sample run
(s), the component concentrations of the
sample are calculated as follows:

C

s

= C

c

Amount

s

/

Amount

c