7 Methods

IC Net 2.2

178

parameter is used to detect an underflow
condition.

Zero

A signal level on a baseline (in ADC con-
version units). This parameter is a result of
the ADC calibration.

Maximum

Maximum value of the linear range of the
AD interface (in ADC conversion units). This
parameter is used to detect overflow condi-
tion.

Range

An input signal value in maximum in user-
defined units.

Range

= (

Maximum

-

Zero

)

Coef

Coefficient

ADC sensitivity coefficient (weight of ADC
bit in user-defined units).

Noise

Estimated baseline noise value of the
channel in ADC conversion units (bits).

Shift

Shift of the current channel in time. This pa-
rameter is useful in the case of multi-
channel chromatograms only.

(multi)Channel ratio

Spectral ratios. The option is valid for multi-
channel chromatograms only.