Metrohm IC Net 2.2 User Manual

Page 189

Advertising
background image

7.6 Report output

IC Net 2.2

181

Concentration of std

Concentration of the internal standard

component for relative concentration calcu-
lations. This value is stored in

This run

col-

umn of the concentration table.

Total % for normalization

A value to which the sum of concentrations
is normalized. It is used for

Response nor-

malization

and

Normalized concentrations

quantification methods. The default value is
100.

Printing order

Determines the order of the components in
the peak table.

By peaks

Lists run results for all peaks detected. Uni-
dentified peaks (corresponding to the uni-
versal component) are included, but miss-
ing components are not listed.

By components

Missing components are always included
into the report, even those with concentra-
tion equal to zero. In the case of presence
of universal component the summary for all
unidentified peaks is presented as a single
line.

<Customize>

Calls up a box with a list of columns that
can be included into the report (see section
7.6.8
). It is available only if

Quantification

method = Custom

has been selected.

Report all peaks

Checkbox that includes all peaks irrespec-
tive of the concentration of the substance.
Otherwise the report generator excludes
lines with zero concentration of the compo-
nent. It is available only if

Quantification

method = Custom

has been selected.

Groups

If this checkbox is checked, a separate
peak table report is put out for every group
specified in the component table.

Advertising