2 start sample queue, 3 pause sample queue – Metrohm IC Net 2.2 User Manual

Page 219

Advertising
background image

9.2 Sample queue control

IC Net 2.2

211

Amount

Sample amount. If this value is different for
the calibration run (c) and the sample run
(s), the component concentrations of the
sample are calculated as follows:

C

s

= C

c

Amount

s

/

Amount

c

Shut down system after the queue finishes

If this option is enabled, the system hard-
ware is shut down after the sample queue
has been finished (all pumps are switched
off), but it remains connected.

Close queue dialog window after the queue finishes

If this option is checked, the sample queue
overview table is closed at the end of the
last determination of the queue.

<Start>

Start execution of the sample queue from
the first line with

Started = 0

.

<Pause>

Stop execution of the sample queue after
the current determination has been fin-
ished.

<Edit>

Open the sample queue editor program for
editing the sample queue table (see section
9.3
).

<Close>

Close the sample queue overview table.


9.2.2

Start sample queue

SAMPLE QUEUE OVERVIEW /

Control / Start

Start execution of the sample queue from the first row with

Started

= 0

. For each row, the sample-specific data are transferred to the

method of the current determination and the parameter

Started

is

set to

1

for this row.

A running sample queue can be interrupted with

<Pause>

. It can

be restarted with

<Start>

(with a running chromatogram only when

data acquisition is finished).

9.2.3

Pause sample queue

SAMPLE QUEUE OVERVIEW /

Control / Pause

Interrupt execution of the sample queue after the current determi-
nation has been finished.
A running sample queue can be interrupted with

<Pause>

. When

the current determination has been finished, the sample queue
can be restarted with

<Start>

or stopped with

Control / Stop deter-

mination

in the

SYSTEM

window.

Advertising