7 Methods

IC Net 2.2

166

7.5.5 Peak

identification

IC NET / Method / Calibration / Identification

Selecting this menu item or clicking on

<Identification>

in the

component table opens the

Peak identification

window with pa-

rameters for tuning the peak identification procedure.

Number of components

Number of components in the component

table (read-only).

Scheme

Quick choice of identification parameters
scheme.

Standard

Set default meanings of identification pa-
rameters:

Height

for reference components,

Time

for all other components (ordinary

components).

Nonstandard

Set custom meanings of identification
parameters.

Identification

Identification

parameters.

Reference peaks

Identification parameter for reference com-
ponents. Default is

Height

.

Other peaks

Identification parameter for ordinary com-
ponents. Default is

Time

.

Retention units

Choice of retention units. Default is

min

.

Retention time

Retention

time.

<Update>

Updates expected retention times of com-
ponents according to the current chroma-
togram.