Metrohm IC Net 2.2 User Manual

Page 71

Advertising
background image

6.4 761 Compact IC

IC Net 2.2

63

Conductivity,

µ

S/cm

Live display of measured conductivity.

Pressure, MPa

Live display of measured pressure.
The color settings for this two fields can be
changed by clicking the fields with the right
mouse button (see section 6.2.2).

Actual values

Display of actual values.

Flow, mL/min

Display of current flow rate of the high-
pressure pump.

Pressure max, MPa

Display of current maximum pressure limit
for high-pressure pump.

Pressure min, MPa

Display of current minimum pressure limit
for high-pressure pump.

Full scale,

µ

S/cm

Display of current full scale range.

Remote lines

Display of current remote line settings.

Valve

Injection

valve.

<Inject>

Switch injection valve to "INJECT" position.

<Fill>

Switch injection valve to "FILL" position

Suppressor

Suppressor

module.

<Step>

Switch the suppressor module to the next
position. The time since the last switching
of the suppressor module is displayed in
the field beside the

<Step>

button.

IC pump

High-pressure

pump.

<On>

Start high-pressure pump.

<Off>

Stop high-pressure pump.

Peristaltic

Peristaltic

pump.

<On>

Start peristaltic pump.

<Off>

Stop peristaltic pump.

System startup values

The system startup values are sent and ap-

plied to the 761 Compact IC each time the
system is connected, a determination is
started, or the values are sent manually with

<Send to unit>

.

Flow, mL/min

Startup value for flow rate of the high-
pressure pump.
Entry range:

0.20

...

2.50 mL/min

Pressure max, MPa

Startup value for maximum pressure limit
for high-pressure pump. This limit is con-
trolled even without connection to the PC.
Entry range:

0.0

...

25.0 MPa

Pressure min, MPa

Startup value for minimum pressure limit for
high-pressure pump. This limit is controlled