5.2 762 IC Interface

IC Net 2.2

39

5.2.4

762 IC Interface window

762 icon / Open

The

762 IC Interface

window for parameter settings is opened by

selecting this menu option with the right mouse button or by click-
ing the

762 icon

on the toolbar. It consists of three tabs

Init

,

RS set-

tings

, and

Links

.

Init

The

Init

tab of the

762 IC Interface

window contains data acquisi-

tion parameters for the 762 IC Interface.

Label

Optional label to name the interface with
maximum 8 characters.

ROM version

Instrument program version number.

Measuring per second

Number of data points measured per s.

Entry range:

10, 20, 30, 50, 60 points/s

Range

Range for AD converter.
Example: converts

±

2500 mV to 2

24

bits

Entry range:

2500, 1250, 625, 312.5, 156.25,

78.125, 39.062 mV

Second press on RUN/STOP means 'stop' when measuring

If this option is enabled, the running data
acquisition will be stopped if the RUN/STOP
button on the 762 IC Interface front panel is
pressed. This is equal to

Stop determination

.

If this option is disabled, the running data
acquisition will not be stopped if the
RUN/STOP button on the 762 IC Interface
front panel is pressed.