7.5 Calibration and quantification

IC Net 2.2

163

7.5.3

External standard calibration

External standard calibration (absolute calibration) is a key way of
calibration. Basically this procedure calculates the dependence of
injected component's quantity versus area (or height) of the cor-
responding peak. This dependence is showed as a calibration
plot with injected quantity along the Y axis and peak area (or
height) along the X axis. The injected quantity Q

i

is calculated as

product of component's concentration C

i

to the reduced injected

sample volume V':

Q

i

= C

i

V'

The calibration function W

i

(R)

for each component is calculated

using the RSD method.

W

i

(R) = RSD(Q

ij

, R

ij

) = RSD((C

i

V'), R

ij

)

When quantification procedure is performed, a concentration C

i

(absolute concentration of a component in the sample) is deter-
mined as a ratio of the computed quantity W

i

(R

i

)

to the corrected

volume of sample injected V':

C

i

= W

i

(R

i

) / V'

7.5.4 Component

table

IC NET / Method / Calibration / Components

If the component table is opened, the chromatogram window is
split into two parts. The upper one shows the chromatogram, the
component table appears in the bottom part of the screen. When
moving into the component table, a special cursor in the upper
part of the window jumps to the peak corresponding to the cur-
rent component.

The component table stores the following information on the
components to be analyzed: