7.6 Report output

IC Net 2.2

183

7.6.8 Report

elements

If

Quantification method = Custom

has been selected, the peak ta-

ble can be customized on the user's choice. After clicking the

<Customize>

button, the following columns can be included in the

peak table.

number

Peak

number.

retention time

Retention time of the component (in min-
utes, irrespective of the chosen retention
units on the chromatogram graph axes).
The total value in the column is equal to the
chromatogram duration.

halfwidth

Width of the peak at half height (in min-
utes).

height

Height of the peak (in mV). The total value
in this column is the sum of heights for all
identified peaks.

height%

Normalization of peak heights for all peaks
using the NORM value entered in the

Total

% for normalization

field (default setting

100%):

H

i

% = NORM

H

i

/

H

i

area

Area of the peak. Depends on the units on
the X and Y axes of the chromatogram. The
total value in this column is the sum of ar-
eas for all identified peaks (including uni-
versal component).

area%

Normalization of peak areas for all peaks
using the NORM value entered in the

Total

% for normalization

field (default setting

100%):

A

i

% = NORM

A

i

/

A

i

capacity factor

The capacity factor k'

i

of the component is

equal to the ratio of its corrected retention
time (t - t

0

)

to the void time of the system t

0

:

k'

i

= (t

i

- t

0

) / t

0

Total for this column is equal to the capac-
ity factor of the last peak of the chroma-
togram.

resolution

Resolution R for two neighboring peaks is
calculated as:

R = (t

i+1

- t

i

) / (w

0.607i

+ w

0.607(i+1)

)